alivecor kardiamobile_firmware 敏感数据加密缺失
CVE编号
CVE-2022-41627利用情况
暂无补丁情况
N/A披露时间
2022-10-28漏洞描述
The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.解决建议
建议您更新当前系统或软件至最新版,完成漏洞的修复。
参考链接 |
|
---|---|
https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01 |
受影响软件情况
# | 类型 | 厂商 | 产品 | 版本 | 影响面 | ||||
1 | |||||||||
---|---|---|---|---|---|---|---|---|---|
运行在以下环境 | |||||||||
系统 | alivecor | kardiamobile_6l_firmware | - | - | |||||
运行在以下环境 | |||||||||
系统 | alivecor | kardiamobile_card_firmware | - | - | |||||
运行在以下环境 | |||||||||
系统 | alivecor | kardiamobile_firmware | - | - | |||||
运行在以下环境 | |||||||||
硬件 | alivecor | kardiamobile | - | - | |||||
运行在以下环境 | |||||||||
硬件 | alivecor | kardiamobile_6l | - | - | |||||
运行在以下环境 | |||||||||
硬件 | alivecor | kardiamobile_card | - | - |
- 攻击路径 相邻
- 攻击复杂度 低
- 权限要求 无
- 影响范围 未更改
- 用户交互 无
- 可用性 高
- 保密性 低
- 完整性 低
CWE-ID | 漏洞类型 |
CWE-311 | 敏感数据加密缺失 |
CWE-319 | 敏感数据的明文传输 |